Diagnostic Accuracy of the Rapid Xpert HIV-1 Viral Load XC, Xpert HIV-1 Viral Load, & m-PIMA HIV-1/2 Viral Load in South African Clinics

6Citations signalées — pas une note de qualité
6Institutions déclarées
3Pays d’affiliation déclarés

Résumé fourni par la source

BACKGROUND: We aimed to evaluate the analytic performance of 3 rapid HIV viral load assays: the novel Xpert HIV-1 VL XC (Xpert XC), Xpert HIV-1 VL (Xpert VL), and m-PIMA HIV-1/2 VL (m-PIMA). SETTING: Two South African clinics. METHODS: We conducted a prospective diagnostic accuracy study. Site-laboratory technicians and nurses used the Xpert XC, Xpert VL, and m-PIMA to test plasma samples from people with HIV receiving antiretroviral therapy. We compared results with the Roche cobas HIV-1 reference assay. We determined accuracy to detect viraemia at the World Health Organization (WHO) failure threshold of 1000 copies/mL on all 3 assays, and 50 and 200 copies/mL on the Xpert assays. We assessed the agreement using Bland-Altman plots. RESULTS: We enrolled 140 participants (98 [70%] women, median age 37 years), who provided 189 paired samples at one or more timepoints. We tested 174 samples with the Xpert XC, 188 with the Xpert VL, and 128 with the m-PIMA. At 1000 copies/mL, sensitivity and specificity (95% confidence intervals) were 97% (82 to 100) and 98% (93 to 99) (Xpert XC), 100% (87 to 100) and 96% (91 to 98) (Xpert VL), and 92% (72 to 99) and 99% (93 to 100) (m-PIMA) respectively. At 50 copies/mL, sensitivity and specificity were 93% (81 to 98) and 96% (91 to 99) (Xpert XC), and 95% (84 to 99) and 95% (90 to 98) (Xpert VL) respectively. Mean bias was -0.10 (-0.54 to 0.34) log10 copies/mL (Xpert XC), 0.07 (-0.37 to 0.52) log10 copies/mL (Xpert VL), and -0.26 (-0.83 to 0.31) log10 copies/mL (m-PIMA). CONCLUSIONS: In these South African clinics, the accuracy of all 3 assays was clinically acceptable to detect viraemia at the WHO failure threshold, whereas both Xpert assays were also accurate at detecting low-level viraemia.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.

Titre Crossref
Diagnostic Accuracy of the Rapid Xpert HIV-1 Viral Load XC, Xpert HIV-1 Viral Load, & m-PIMA HIV-1/2 Viral Load in South African Clinics
Date Crossref
13/06/2022
Éditeur
Ovid Technologies (Wolters Kluwer Health)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

HIV Research and TreatmentHIV/AIDS drug development and treatmentHIV/AIDS Research and Interventions

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, ROR et la Banque mondiale, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune donnée externe enregistrée en base. Sources et limites.