Use of Portable Air Purification Systems to Eliminate Aerosol Particles from Patient Rooms

3Citations signalées — pas une note de qualité
3Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

SARS-CoV-2 is mainly transmitted via airborne aerosols. We determined the efficacy of two different indoor air-purification systems and compared them to conventional measures of room ventilation. Radioactive particles with a mass median aerodynamic diameter of 0.6 ± 0.4 µm were nebulized from the head position of a simulated patient inside a 38 m3 measuring room. Air samples were drawn using an artificial lung from outside the room from the head and belly position of the simulated patient. The radioactivity from these air samples was determined in % of the minute-by-minute release of aerosol by the simulated patient. The samples were taken with the windows closed, with the windows open and with two different air purification systems in different locations. Opening the windows while the fan was working did not reduce the cumulative mass of sampled particles during ongoing nebulization. Only the more powerful air purifier was able to decrease cumulative sampled radioactivity almost to zero when positioned close to the emitting particle source. We observed air-turbulences caused by the air intake of the air purifier resulting in regional particle concentration peaks. Measurements and calculations demonstrate that indoor air purification systems can be effective measures to eliminate airborne particles. The air cleaning capacity of the system must be matched to the room size and the air intake of the units should be positioned close to the particle source.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Use of Portable Air Purification Systems to Eliminate Aerosol Particles from Patient Rooms
Date Crossref
01/01/2022
Éditeur
Springer Science and Business Media LLC
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Infection Control and VentilationFire dynamics and safety research

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.