Assessment of two POC technologies for CD4 count in Morocco

0Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : Côte d’Ivoire. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Abstract Background : In the era of “test and treat strategy”, CD4 testing remains an important tool for monitoring HIV-infected individuals. Since conventional methods of CD4 count measurement are costly and cumbersome, POC CD4 counting technique are more affordable and practical for countries with limited resources. Before introducing such methods in Morocco, we decided to assess their reliability. Methods : in this study 92 blood samples from HIV-infected patients, were tested by PIMA to generate absolute CD4 count and by FACSPresto to derive percentage and absolute CD4 count. Flow cytometry using FacsCalibur, was used as reference method for CD4 count comparison. Linear regression, Bland and Altman analysis were performed to assess correlation and agreement between POC methods and the reference method. In addition, sensibility and specificity, positive predictive value (PPV), negative predictive value (NPV) at 350 CD4 count threshold were also determined. Finally, because FACSPresto can also measure hemoglobin (Hb) concentration, 52 samples were used to compare FACSPresto against an automated hematology analyzer. Results : With PIMA technology, the coefficient of determination R 2 was 0.93. Regarding FACSPresto, R 2 was 0.93 and 0.96for absolute CD4 count and percentage CD4 count, respectively. Bland and Altman analysis displayed a mean bias of -32.3 cells/µl with a limit of agreement (LOA): -181.3 to116.8, for Pima. The mean bias for FACSPresto was -8.1 cells/µl with LOA:-158.2 to 142.0for absolute CD4 count and 0.2 with LOA:-3.8 to 4.3, for CD4 percentage. Moreover, with a threshold of 350 CD4 count, sensibility, specificity, PPV, NPV, were 98%, 87%, 89%, 86%, and 88%, 96%,93% and 92% for PIMA and FACSPresto, respectively. Finally, the hemoglobin measurement evaluation displayed an R 2 of 0.80 and a mean bias of-0.12 with a LOA between -1.75 and 1.51. Conclusion : When compared to the reference method, PIMA and FACSPresto have shown good performance, for CD4 counting. The introduction of such POC technology will speed up the uptake of patients in the continuum of HIV care, in our country.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Assessment of two POC technologies for CD4 count in Morocco
Date Crossref
22/01/2020
Éditeur
Research Square Platform LLC
Type
posted-content

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Blood groups and transfusionTuberculosis Research and EpidemiologyHIV Research and Treatment

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.