O08 BIOMECHANICS OF MANUAL, STAPLED, AND MANUAL-STAPLED SUTURES ON FASCIA LATA. TIPS FOR ABDOMINAL WALL SURGEONS

0Citations signalées, ce qui n’est pas une note de qualité
4Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : it. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Abstract Aim No data on the biomechanical properties of staplers’ use in the repair of abdominal wall defects are available. Our objective is to study prospectively the biomechanical properties of manual, stapled and mixed manual/stapled fascial sutures. Material-Methods Stress/strain tests were performed on 16 human cadaver fascial samples. The data on strength, strain (deformability), Young’s modulus (elasticity), and dissipated energy (toughness modulus) were recorded for each type of suture. Results Stapled and mixed suture showed a significantly higher strength (manual 0.83, stapled 2.10, mixed 2.68 MPa), and a trend towards a higher strain as compared to manual sutures (manual 344, stapled 249, mixed 280%). Stapled and mixed suture were four-fold higher as compared to manual sutures (manual 1.779, stapled 7.374, mixed 6.964 MPa). Manual and mixed sutures showed significantly higher dissipated energy (manual 0.99, vs stapled 0.73, vs mixed 1.35 mJ-mm3). Conclusions Stapled and mixed sutures have better strength performances than manual sutures. On the other hand, stapled and mixed sutures have significantly higher Young’s modulus and lower ultimate strain, showing less deformability, possibly translating in less efficiency in large midline defects, where manual sutures might display higher elasticity. Also, hand-sewn sutures (in manual or over-sewn stapled) appear to increase the ability to absorb mechanical energy, whereas stapled sutures showed to be less tough. Furthermore, an over-sewn stapled suture, as compared to a stapled suture, gains in term of strength, ultimate strain, Young’s modulus, and dissipated specific energy, suggesting the need of over-sewn stapled sutures in case of larger fascial defects.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
O08 BIOMECHANICS OF MANUAL, STAPLED, AND MANUAL-STAPLED SUTURES ON FASCIA LATA. TIPS FOR ABDOMINAL WALL SURGEONS
Date Crossref
01/11/2021
Éditeur
Oxford University Press (OUP)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Surgical Sutures and AdhesivesAnatomy and Medical TechnologyHernia repair and management

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.