Surrogate test performance for SARS‐CoV ‐2 neutralizing antibodies ( nAbs ) for convalescent plasma ( CCP ): How useful could they be?

7Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : br. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: COVID-19 high-titer CCP selection is a concern, because neutralizing antibody (nAb) testing requires sophisticated labs and methods. Surrogate tests are an alternative for measuring nAb levels in plasma bags, including those that are pathogen-reduced. STUDY DESIGN/METHODS: We studied a panel consisting of 191 samples from convalescent donors tested by nAb (CPE-VNT), obtained from 180 CCP donations (collection: March 20-January 21) and 11 negative controls, with a total of 80 and 111 serum and plasma samples (71 amotosalen/UV treated), with nAb titers ranging from negative to 10,240. Samples were blindly tested for several surrogates: one anti-RBD, two anti-spike, and four anti-nucleocapsid tests, either isolated or combined to improve their positive predictive values as predictors of the presence of high-titer nAbs, defined as those with titers ≥160. RESULTS: Except for combined and anti-IgA/M tests, all isolated surrogate tests showed excellent performance for nAb detection: sensitivity (98.3%-100%), specificity (85.7%-100%), PPV (98.9%-100%), NPV (81.3%-100%), and AUC (0.93-0.96), with a variable decrease in sensitivity and considerably lower specificity when using FDA authorization and concomitant nAb titers ≥160. All surrogates had AUCs that were statistically different from CPE-VNT if nAb≥160, including when using combined, orthogonal approaches. CONCLUSIONS: Surrogate tests (isolated or in combination) have an indirect good performance in detecting the presence of nAb, with lower sensitivity and specificity when high nAb titer samples are used, possibly accepting a considerable number of donors whose nAb titers are actually low, which should be evaluated by each laboratory responsible for CCP collection.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.

Titre Crossref
Surrogate test performance for <scp>SARS‐CoV</scp>‐2 neutralizing antibodies (<scp>nAbs</scp>) for convalescent plasma (<scp>CCP</scp>): How useful could they be?
Date Crossref
27/10/2021
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

SARS-CoV-2 and COVID-19 ResearchSARS-CoV-2 detection and testingMonoclonal and Polyclonal Antibodies Research

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.