Laboratory Calibration of the Spectrum Field Scout TDR 300
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Soil moisture sensors (SMSs) are a useful tool that aid in data-driven water management decisions. However, default factory calibrations can be inaccurate and soil-specific calibrations are often required to obtain higher accuracy in the determination of soil water storage and plant available water. In this study, we conducted a lab calibration for the Field Scout TDR 300, which is a popular SMS used in the turfgrass industry. Five soils of different soil textural classes were packed in containers with known soil moisture for the laboratory calibration. The logarithmic model best fit the data for the course- and fine-textured soils, with a root mean square error (RMSE) value of 0.027 and 0.035 cm3 cm-3, respectively. These two calibration curves help to estimate volumetric water content more accurately for native and sand-based soils.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Laboratory Calibration of the Spectrum Field Scout TDR 300
- Date Crossref
- 01/01/2021
- Éditeur
- Kansas State University
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.