The performance of point-of-care antibody test for COVID-19 diagnosis in a tertiary hospital in Bandung, Indonesia
Rattachement africain : id. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
INTRODUCTION: We aim to describe the performance of combined IgM and IgG point-of-care antibody test (POC-Ab) (Wondfo®) compared to real-time reverse transcriptase (rRT-PCR) (Allplex™ 2019-nCoV Assay) in detecting coronavirus disease 2019 (COVID-19). METHODOLOGY: We compared POC-Ab with rRT-PCR results among patients in a tertiary hospital from January to March 2020 in Bandung, Indonesia. We selected presumptive COVID-19 patients with positive rRT-PCR consecutively and 20 patients with negative rRT-PCR results were selected randomly from the same group of patients as controls. We described the sensitivity, specificity, positive predictive value (PPV), negative predictive value (NPV) with corresponding 95% confidence interval using serum and capillary blood samples. We also tested POC-Ab using non-COVID-19 (confirmed dengue and typhoid) patients' sera. RESULTS: Twenty-seven patients with positive rRT-PCR result and 20 negative controls were included (68.1% males, mean age 46 (SD: 15.4)). Using the serum, the sensitivity of the POC-Ab was 63.0% (42.4-80.6), specificity was 95.0% (75.1-99.9), PPV was 94.4% (72.7-99.8), NPV was 65.5% (45.7-82.1). A subset of 20 patients was tested using a capillary blood sample. The accuracy of the capillary blood sample is lower compared to serum (50.0% vs. 78.7%). None of the non-COVID-19 sera tested were reactive. CONCLUSIONS: POC-Ab for COVID-19 has a high specificity with no false-positive result in non-COVID-19 sera. Therefore, it can be used to guide diagnostic among symptomatic patients in resource limited settings. Given its low sensitivity, patients with high suspicion of COVID-19 but non-reactive result should be prioritized for rRT-PCR testing.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- The performance of point-of-care antibody test for COVID-19 diagnosis in a tertiary hospital in Bandung, Indonesia
- Date Crossref
- 07/03/2021
- Éditeur
- Journal of Infection in Developing Countries
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.