A novel wireless implant for central venous pressure measurement: First animal experience

3Citations signalées, ce qui n’est pas une note de qualité
3Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : ie, us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background/ObjectiveCentral venous pressure (CVP) serves as a surrogate for right atrial pressure, and thus could potentially predict a wider range of heart failure conditions. The purpose of this work is to assess CVP, through an implantable sensor incorporated with a novel anchor design, in the inferior and superior vena cava of an animal model.MethodsTwo animals (Dorset sheep) were implanted with sensors at 3 different locations: inferior vena cava (IVC), superior vena cava (SVC), and pulmonary artery (PA). Two sensors with distinct anchor designs considering anatomical requirements were used. A standard PA sensor (trade name Cordella) was deployed in the PA and SVC, whereas a sensor with a modified cylindrical anchor with various struts was designed to reside in the IVC. Each implant was calibrated against a Millar catheter reference sensor. The ability of the central venous sensors to detect changes in pressure was evaluated by modifying the fluid volume of the animal.ResultsThe sensors implanted in both sheep were successful, which provided an opportunity to understand the relationship between PA and CVP via simultaneous readings. The mapping and implantation in the IVC took less than 15 minutes. Multiple readings were taken at each implant location using a hand-held reader device under various conditions. CVP recorded in the IVC (6.49 mm Hg) and SVC (6.14 mm Hg) were nearly the same. PA pressure (13–14 mm Hg) measured was higher than CVP, as expected. The SVC waveforms showed clear beats and respiration. Respiration could be seen in the IVC waveforms, but not all beats were easily distinguishable. Both SVC and IVC readings showed increases in pressure (3.7 and 2.7 mm Hg for SVC and IVC, respectively) after fluid overload was induced via extra saline administration.ConclusionIn this work, the feasibility of measuring CVP noninvasively was demonstrated. The established ability of wireless PA pressure sensors to enable prevention of decompensation events weeks ahead can now be explored using central venous versions of such sensors. Central venous pressure (CVP) serves as a surrogate for right atrial pressure, and thus could potentially predict a wider range of heart failure conditions. The purpose of this work is to assess CVP, through an implantable sensor incorporated with a novel anchor design, in the inferior and superior vena cava of an animal model. Two animals (Dorset sheep) were implanted with sensors at 3 different locations: inferior vena cava (IVC), superior vena cava (SVC), and pulmonary artery (PA). Two sensors with distinct anchor designs considering anatomical requirements were used. A standard PA sensor (trade name Cordella) was deployed in the PA and SVC, whereas a sensor with a modified cylindrical anchor with various struts was designed to reside in the IVC. Each implant was calibrated against a Millar catheter reference sensor. The ability of the central venous sensors to detect changes in pressure was evaluated by modifying the fluid volume of the animal. The sensors implanted in both sheep were successful, which provided an opportunity to understand the relationship between PA and CVP via simultaneous readings. The mapping and implantation in the IVC took less than 15 minutes. Multiple readings were taken at each implant location using a hand-held reader device under various conditions. CVP recorded in the IVC (6.49 mm Hg) and SVC (6.14 mm Hg) were nearly the same. PA pressure (13–14 mm Hg) measured was higher than CVP, as expected. The SVC waveforms showed clear beats and respiration. Respiration could be seen in the IVC waveforms, but not all beats were easily distinguishable. Both SVC and IVC readings showed increases in pressure (3.7 and 2.7 mm Hg for SVC and IVC, respectively) after fluid overload was induced via extra saline administration. In this work, the feasibility of measuring CVP noninvasively was demonstrated. The established ability of wireless PA pressure sensors to enable prevention of decompensation events weeks ahead can now be explored using central venous versions of such sensors.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
A novel wireless implant for central venous pressure measurement: First animal experience
Date Crossref
01/11/2020
Éditeur
Elsevier BV
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Mechanical Circulatory Support DevicesHemodynamic Monitoring and TherapyCardiac pacing and defibrillation studies

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.