Evaluation of the Alinity m HIV-1 assay for the quantification of HIV-1 RNA plasma viral load in a high-throughput molecular laboratory in South Africa
Résumé fourni par la source
Regular HIV-1 viral load monitoring forms an essential part of any successful HIV-1 treatment programme. Abbott Molecular recently released the Alinity m HIV-1 assay to be run on the Alinity m System, a fully automated, continuous and random access analyser using ReadiFlex™ technology. Our study investigated the performance of the Alinity m HIV-1 assay in comparison to the cobas® HIV-1 test in a high-throughput molecular laboratory. We compared the performance of the Alinity m HIV-1 assay with the cobas® HIV-1 test, performed on both the cobas® 4800 and cobas® 6800 systems at three clinically relevant thresholds (50, 200 and 1000 cp/mL). Excellent correlation (r = 0.98) and agreement (mean bias −0.004 Log10 cp/mL) was achieved between the cobas® 4800 and Alinity m HIV-1 assay. While there was good correlation between the Alinity m HIV-1 assay and the cobas® 6800 (r = 0.99), Bland-Altman analysis indicated that the cobas® 6800 on average measured 0.22 Log10 cp/mL higher than the Alinity m HIV-1 assay across the dynamic range. Percentage agreement was excellent at the 200 cp/mL and 1000 cp/mL thresholds and was slightly lower at 50 cp/mL in comparison with the cobas® systems. The Alinity m HIV-1 assay compared well with the cobas® HIV-1 test on both the cobas® 4800 and cobas® 6800 systems in a high-throughput molecular laboratory in South Africa, a low- to middle-income country.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Evaluation of the Alinity m HIV-1 assay for the quantification of HIV-1 RNA plasma viral load in a high-throughput molecular laboratory in South Africa
- Date Crossref
- 01/11/2020
- Éditeur
- Elsevier BV
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.
Institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.