Third-Generation Balloon and Self-Expandable Valves for Aortic Stenosis in Large and Extra-Large Aortic Annuli From the TAVR-LARGE Registry

44Citations signalées, ce qui n’est pas une note de qualité
18Institutions déclarées
6Pays d’affiliation déclarés

Rattachement africain : es, us, nl, ca, ch, fr. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Background: Currently, 2 third-generation transcatheter valves, 29-mm Sapien-3 and 34-mm Evolut-R (ER), are indicated for large sized aortic annuli. We analyzed short and 1-year performance of these valves in patients with large (area ≥575 mm 2 or perimeter ≥85 mm) and extra-large (≥683 mm 2 or ≥94.2 mm) aortic annuli undergoing transcatheter aortic valve replacement. Methods: A total of 833 patients across 12 centers with symptomatic aortic stenosis and large aortic annuli underwent transcatheter aortic valve replacement with 29-mm Sapien-3 (n=640) or 34-mm ER (n=193). Clinical, anatomic, and procedural characteristics were collected, and Valve Academic Research Consortium-2 outcomes were reported. Results: Median aortic annulus area and perimeter were 617 mm 2 (591–657) and 89.1 mm (87.0–92.1), respectively (704 mm 2 [689–743] and 96.0 mm [94.5–97.9] in the subgroup of 124 patients with extra-large annuli). Overall device success was 94.3% (Sapien-3, 95.8% and ER, 89.3%; P =0.001), with a higher rate of significant paravalvular leak ( P =0.004), second valve implantation ( P =0.013), and valve embolization ( P =0.009) in the ER group. Thirty-day and 1-year mortality was 2.4% and 9.2%, respectively, without differences between groups. Valve hemodynamics were excellent (mean gradient, 8.8±3.6 mm Hg; 3.3% rate of moderate-severe paravalvular leak) in the extra-large annulus, without differences compared with the large annulus group. Conclusions: In patients with large and extra-large aortic annuli, transcatheter aortic valve replacement using 29-mm Sapien-3 and 34-mm ER is safe and feasible. Observed differences in clinical outcomes and hemodynamic performance may guide valve choice in this cohort of patients undergoing transcatheter aortic valve replacement.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Third-Generation Balloon and Self-Expandable Valves for Aortic Stenosis in Large and Extra-Large Aortic Annuli From the TAVR-LARGE Registry
Date Crossref
01/08/2020
Éditeur
Ovid Technologies (Wolters Kluwer Health)
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Cardiac Valve Diseases and TreatmentsAortic Disease and Treatment ApproachesCongenital Heart Disease Studies

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.