Reliability of Spectral-Domain OCT Ellipsoid Zone Area and Shape Measurements in Retinitis Pigmentosa
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Purpose: We investigate the ellipsoid zone (EZ) area and EZ boundary shape measurement reliability and the operability characteristics of two methods of EZ boundary delineation in spectral-domain optical coherence tomography (SD-OCT). Methods: EZ boundaries in SD-OCT scans of 122 eyes from 64 subjects with autosomal dominant retinitis pigmentosa were delineated by three raters using two methods, termed the profile and en face methods. For each method, we determined the measurement reliabilities for boundary area (EZ area) and boundary shape, percentage of eyes with measurable EZ (measurability), time required, and effect of rater experience. Results: With expert raters, inter- and intrarater area intraclass correlation coefficients (ICCs) were 0.986 and 0.980 (profile) and 0.959 and 0.976 (en face), respectively. In comparison, the corresponding shape ICCs were 0.906 and 0.891 (profile) and 0.845 and 0.885 (en face), indicating lower reliability for the raw measurements (P ≤ 0.01). Only profile method interrater reliability depended on experience. Average measurement times per eye were 8.2 (profile) and 4.1 (en face) minutes. Measurability percentages were 99.2% (profile) and 73.0% (en face). Conclusions: The slower profile method had better measurability, and with expert raters yielded the best area and shape reliabilities. The faster, but less sensitive, en face method still showed excellent reliability, and was less dependent on experience. Shape analysis reveals the boundary measurements underpinning EZ area have lower reliability than suggested by area analysis. Translational Relevance: This study provides new reliability perspectives and logistical considerations for the manual measurement procedures that generate EZ area outcome measures.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Reliability of Spectral-Domain OCT Ellipsoid Zone Area and Shape Measurements in Retinitis Pigmentosa
- Date Crossref
- 11/06/2019
- Éditeur
- Association for Research in Vision and Ophthalmology (ARVO)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.