Benchmarking the DACS ‐integrated Radiation Dose Monitor® skin dose mapping software using XR ‐ RV 3 Gafchromic® films

20Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : fr. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Purpose To perform a benchmark of a new DACS‐integrated patient skin dose mapping solution using on‐phantom measurements with Gafchromic® films. Materials and methods To calculate cumulative patient skin dose distribution with 1‐cm2 resolution, a Radiation Dose Monitor (RDM, Medsquare), using the Radiation Dose Structured Report (RDSR), tabulated backscatter and mass energy absorption coefficients together with site‐specific corrections for table, mattress attenuation, and air kerma calibration factor. Peak skin dose (PSD) and two‐dimensional (2D) skin dose distributions calculated with RDM were compared against on‐phantom measurements with XR‐RV3 Gafchromic® films considering two widely used x‐ray equipment. Seventeen different settings which include simple and multiple beam projections with extreme angulations (up to 75°), all available fields‐of‐view (FOVs 48–11 cm), additional collimation, variable table height and lateral positions, and variable phantom thickness (12, 20, and 30 cm) were involved. Results Due to a careful calibration of films using clinical beam qualities, 22.8% (k = 2) overall measurement uncertainty was achieved. Calculated and measured PSD values agreed with an average difference of 10% ± 7% and 9% ± 7% for 34 test conditions performed on Siemens Artis Zee and GEMS Innova IGS interventional systems, respectively. Finally, RDM's 2D skin dose maps closely matched those registered on XR‐RV3 films considering the 1‐cm2 resolution. While RDM correctly reproduced beam overlapping due to variable tube projections, FOV, table positions, etc., few challenges were identified related to conversion of rectangular fields to square areas in the RDSR and a stair‐step effect visible for large tube projections (>45°). Conclusion The accuracy of RDM's DACS‐integrated skin dose mapping software was acceptable considering measurement uncertainties associated with Gafchromic® films.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.

Titre Crossref
Benchmarking the <scp>DACS</scp>‐integrated Radiation Dose Monitor® skin dose mapping software using <scp>XR</scp>‐<scp>RV</scp>3 Gafchromic® films
Date Crossref
31/08/2018
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Advanced Radiotherapy TechniquesRadiation Dose and ImagingDigital Radiography and Breast Imaging

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.