P3.168 Evaluation of the point-of-care xpert® CT/NG and osom® trichomonas rapid tests against the anyplex™ii STI-7 detection assay

1Citations signalées, ce qui n’est pas une note de qualité
3Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : Afrique du Sud. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Introduction Syndromic management of sexually transmitted infections (STIs), as practised in most poorly resourced countries misses out asymptomatic infections. Affordable nucleic acid amplification tests (NAATs) are needed for accurate STI diagnosis and treatment. Methods As part of a cohort study assessing a diagnostic STI care model among young South African women presenting for syndromic care, we evaluated the clinic-based point-of-care (POC) tests Xpert CT/NG and OSOM Trichomonas Rapid Test against the laboratory-based Anyplex II STI-7, a multiplex real-time PCR assay which detects Chlamydia trachomatis (CT), Neisseria gonorrhoeae (NG), Trichomonas vaginalis (TV), M. genitalium (MG), M. hominis (MH), U. urealyticum (UU) and U. parvum (UP) in a single reaction. All positive and discordant results were confirmed with a third molecular assay, the FTD STD9. Results Vaginal swabs taken from 247 women were assessed in parallel. 238 of 247 (96.4%) results were in agreement comparing Xpert and Anyplex. All nine discrepant results were positive for CT on Xpert but negative on Anyplex. FTD STD9 confirmed three positive and six negative results. The sensitivity, specificity, positive predictive value (PPV) and negative predictive value (NPV) of Xpert for CT against the two assays was 100%, 97.1%, 86.0%, 100%, respectively; and for NG 100%, 99.6%, 92.3%, 100%. The sensitivity, specificity, PPV and NPV of OSOM for TV against the two assays was 77.8%, 100%, 100%, 99.2%. In addition to the CT, NG and TV detection, the Anyplex identified a prevalence of 4.8% MG, 33.5% MH, 19.1% UU and 51.4% UP in this population. Conclusion The overall performance of Xpert CT/NG against laboratory-based assays was comparable. A lower PPV may lead to some overtreatment, however, in a high burden STI and HIV region, where STIs are often asymptomatic, the high sensitivity and specificity are reassuring. The widened spectrum of Anyplex targets highlights the high burden of Ureaplasma and Mycoplasma species in this setting, whose clinical significance need further exploration.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
P3.168 Evaluation of the point-of-care xpert® CT/NG and osom® trichomonas rapid tests against the anyplex™ii STI-7 detection assay
Date Crossref
01/07/2017
Éditeur
BMJ Publishing Group Ltd
Type
proceedings-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Reproductive tract infections researchCervical Cancer and HPV ResearchSyphilis Diagnosis and Treatment

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.