O10.2 A performance evaluation of the atlas genetics ltd io® system: a novel and rapid point-of-care in vitro diagnostic test forchlamydia trachomatis

0Citations signalées, ce qui n’est pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : gb. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Introduction Rapid Point-Of-Care Tests (POCTs) for Chlamydia trachomatis (CT) may reduce onward transmission and reproductive sexual health (RSH) sequelae by reducing turnaround times between diagnosis and treatment. The io single module system (Atlas Genetics Ltd) runs clinical samples through a microfluidic CT cartridge, delivering results in 30 min. We evaluated its performance in four RSH clinics. Methods 757 females aged >16 provided additional-to-routine self-collected vulvovaginal swab (VVS). Samples were tested fresh on io within 7 days of collection or were frozen at −80°C for later testing. The io CT-assay performance was compared against clinic BD ViperTM Nucleic Acid Amplification Test (NAAT), with discrepant results resolved on the Artus CT/NG assay. The gold standard for discrepants required agreement from 2/3 tests. Factors associated with CT infection were analysed using logistic regression. Results Insufficient volume (n=3), missing clinic NAAT data (n=21), and ‘invalid’ (n=24), where io failed to give a result on two successive runs, meant final analyses were conducted on 709 women (94.3%). CT prevalence was 7.2% (51/709). Sensitivity, specificity, positive (PPV) and negative (NPV) predictive values were respectively: 96.1% (95% Confidence Interval (CI): 86.5–99.5), 97.7% (95% CI: 96.3–98.7), 76.6% (95% CI: 64.3–86.2) and 99.7% (95% CI: 98.9–100). There was no significant difference in performance between fresh and frozen samples, or between symptomatic and asymptomatic patients. Risk factors associated with CT infection were sexual contact CT only. Conclusion The io CT-assay is the only 30 min, fully automated, high-performing NAAT currently CE-marked for CT diagnosis in women, making it a highly promising diagnostic, to enable specific treatment, initiation of partner notification and appropriately intensive health promotion at the point of care.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.

Titre Crossref
O10.2 A performance evaluation of the atlas genetics ltd io® system: a novel and rapid point-of-care in vitro diagnostic test for<i>chlamydia trachomatis</i>
Date Crossref
01/07/2017
Éditeur
BMJ Publishing Group Ltd
Type
proceedings-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Reproductive tract infections researchCervical Cancer and HPV Research

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.