Aller au contenu principal
Accès ouvert déclaré 2000 article

The Muons Registration Efficiency of the CMS ECAL Preshower in the Integrated Test

0Citations signalées, ce qui n’est pas une note de qualité
0Institutions déclarées
0Pays d’affiliation déclarés

Le résumé fourni par la source

The muon registration efficiency of the preshower was estimated from experimental data of the integrated test of the module, that is a prototype of CMS Endcap detector. It is shown that the muon registration efficiency of the preshower plane is 93 % for the muon particles with the energy from 100 to 300 GeV in 3 Tesla axial magnetic field. It was mainly defined by the noise value of the electronics devices AMPLEX SICAL chips. The muons energy loss distribution in the preshower planes was investigated. A mathematical method was proposed to describe the distribution of the energy losses taking into account the intrinsic noise of the electronics devices. A comparison of the experimental data with the simulation results was made.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

Aucun DOI disponible pour le contrôle Crossref.

Les sujets associés

Neutrino Physics ResearchParticle Detector Development and PerformanceMuon and positron interactions and applications

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.