SU‐G‐BRA‐06: Quantification of Tracking Performance of a Multi‐Layer Electronic Portal Imaging Device

0Citations signalées, ce qui n’est pas une note de qualité
4Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

Purpose: The purpose of this study was to quantify the improvement in tumor tracking, with and without fiducial markers, afforded by employing a multi‐layer (MLI) electronic portal imaging device (EPID) over the current state‐of‐the‐art, single‐layer, digital megavolt imager (DMI) architecture. Methods: An ideal observer signal‐to‐noise ratio (d’) approach was used to quantify the ability of an MLI EPID and a current, state‐of‐the‐art DMI EPID to track lung tumors from the treatment beam's‐eye‐view. Using each detector modulation transfer function (MTF) and noise power spectrum (NPS) as inputs, a detection task was employed with object functions describing simple three‐dimensional Cartesian shapes (spheres and cylinders). Marker‐less tumor tracking algorithms often use texture discrimination to differentiate benign and malignant tissue. The performance of such algorithms is simulated by employing a discrimination task for the ideal observer, which measures the ability of a system to differentiate two image quantities. These were defined as the measured textures for benign and malignant lung tissue. Results: The NNPS of the MLI ∼25% of that of the DMI at the expense of decreased MTF at intermediate frequencies (0.25≤ Conclusion: MLI performance in tumor tracking is greatly improved by the additional imager layers. This implies that further improvements in tracking may be gained through increasing the thickness of each MLI layer. For tracking, the MLI performance is limited by noise response. Losses in MTF result in negligible differences in d’. The project was partially supported by a grant from Varian Medical Systems, Inc. and grant No. R01CA188446‐01 from the National Cancer Institute.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
SU‐G‐BRA‐06: Quantification of Tracking Performance of a Multi‐Layer Electronic Portal Imaging Device
Date Crossref
01/06/2016
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Optical Imaging and Spectroscopy TechniquesDigital Radiography and Breast ImagingCalibration and Measurement Techniques

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.