Aller au contenu principal
2013 article

Design of the photomultiplier tube bases for high dynamic range readout in WCDA

7Citations signalées — pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Résumé fourni par la source

The photomultiplier tube (PMT) used in the water Cherenkov detector array (WCDA) of the Large High Altitude Air Shower Observatory (LHAASO) requires a good single photoelectron (SPE) spectrum and a charge dynamic range from 1 to 4000 photoelectrons. In this paper, the bases design and improvement of the photomultiplier tube R5912 are presented. The results show that at the gain of 2.6 × 10 6 , the anode output has a good single photoelectron spectrum, and its charge non-linearity is within 5% when the number of photoelectrons (nPE) is 3500. The charge non-linearity of the 8th dynode output is within 2% when the number of nPE is 4000, which satisfies the dynamic range requirement.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Design of the photomultiplier tube bases for high dynamic range readout in WCDA
Date Crossref
01/03/2013
Éditeur
IOP Publishing
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude et ne compte pas comme une seconde source scientifique indépendante.

Institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Sujets associés

Astrophysics and Cosmic PhenomenaNeutrino Physics ResearchRadiation Detection and Scintillator Technologies

BNTIC News n’est pas le producteur de ces données. Recherche à la demande dans Crossref et Europe PMC, sans clé ; OpenAlex reste optionnel. Aucun service payant requis, aucune réponse conservée. Sources et limites.