Long‐Term Changes in High‐Voltage Impedance of Defibrillating Leads

5Citations signalées, ce qui n’est pas une note de qualité
2Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : us. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: The maturation of the high-voltage impedance (HVLI) of defibrillating leads has not been explored thoroughly. Since impedance influences the charge, current, and energy delivered to the heart in both fixed pulse-width and fixed tilt implantable cardioverter-defibrillator (ICD) waveforms, changes in HVLI might have an effect on the defibrillation threshold. This analysis examined the maturation of defibrillation lead HVLI. METHODS: The data were collected in 515 recipients of ICDs capable of storing high-voltage shock diagnostics, including HVLI. Data with constant superior vena cava (SVC) coil configuration (coil ON or OFF) were collected for up to 24 months. HVLI values were recorded, plotted, and normalized by the value at implant; the percentage of patients in whom HVLI increased or decreased by > or =6 Omega was calculated, and the maturation of leads with SVC ON versus OFF was compared. RESULTS: A > or =6 Omega increase or decrease in HVLI was observed in 41% of patients more than 3 months after ICD implant, with the magnitudes as follows: 6-7.5 Omega change = 16%, 8-9.5 Omega= 11%, 10-11.5 Omega= 6%, and > or =12 Omega= 8%. In this subgroup, 67% of patients showed an increase in impedance, and the remaining 33% of patients showed a decrease. For ICD-programmed SVC OFF, the right ventricular (RV) Coil-to-Can HVLI increased by 8 Omega after 6 weeks and did not significantly change thereafter. CONCLUSIONS: HVLI changed significantly in 41% of leads after 3 months. The HVLI trend of the RV Coil-to-Can configuration appeared to be similar to the SVC Coil/Can-to-RV and Coil-to-Can configurations. Changes in HVLI of 6-12 Omega will alter the characteristics of the shock waveform and could require retuning of the waveforms to maintain adequate DFT safety margin. If re-tuning the shock waveforms is not performed, these HVLI changes have the potential to result in elevated DFT. These HVLI changes may partially account for the 15% increase in DFT over time reported in the literature.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.

Titre Crossref
Long‐Term Changes in High‐Voltage Impedance of Defibrillating Leads
Date Crossref
23/02/2009
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Cardiac pacing and defibrillation studiesCardiovascular Function and Risk FactorsCardiac Valve Diseases and Treatments

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.