Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100
Rattachement africain : nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
BACKGROUND: Determination of leukocyte (WBC) counts in pleural fluid is routinely performed by microscopic examination. In this study, we evaluated the performance of automated (differential) WBC counting in comparison with manual counting. METHODS: Pleural fluid samples (n=45) were obtained from patients undergoing diagnostic thoracocentesis. The manual total WBC count was determined after Samson staining in a Fuchs-Rosenthal hemocytometer; microscopic differential counts were performed on May-Grünwald Giemsa-stained cytospin slides. The Sysmex XE-2100 hematology analyzer was used for automated (differential) WBC counting. The functional detection limit was determined by serial dilution of continuous ambulatory peritoneal dialysis (CAPD) fluid and replicate measurements of each dilution. RESULTS: The automated WBC count (x10(6)/L) was highly correlated with that of the microscopic reference method (r(2)=0.95; WBC-analyzer=0.97 x WBC-reference method+16; n=45). Good agreement was also observed for the absolute lymphocyte count (r(2)=0.92; WBC-analyzer=0.99 x WBC-reference method+32; n=36), neutrophil count (r(2)=0.94; WBC-analyzer=0.91 x WBC-reference method+6; n=35), and monocyte count (r(2)=0.73; WBC-analyzer=0.83 x WBC-reference method+6; n=38). The functional detection limit for WBCs was calculated at 50 x 10(6)/L (coefficient of variation 20%). CONCLUSIONS: With some limitations, total and differential WBC counts in pleural fluid can be reliably determined using the Sysmex XE-2100 instrument.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Automated analysis of pleural fluid total and differential leukocyte counts with the Sysmex XE-2100
- Date Crossref
- 01/01/2006
- Éditeur
- Walter de Gruyter GmbH
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Où se fait cette recherche
-
Erasmus MC pays non établi dans la noticeÉtablissement de santé
-
Erasmus University Rotterdam pays non établi dans la noticeUniversité ou école supérieure
-
Department of Clinical Chemistry pays non établi dans la noticeÉtablissement de santé
Erasmus MC, Erasmus University Rotterdam et Department of Clinical Chemistry.
Une affiliation ne permet pas de déduire la nationalité d’un auteur.