Correction factors for oxygen and flow-rate effects on neonatal Fleisch and Lilly pneumotachometers
Rattachement africain : nl. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
OBJECTIVE: To assess the effects of different oxygen concentrations and flow rates on the measurement errors of neonatal pneumotachometers in heated and unheated situations and to develop correction factors to correct for these effects. DESIGN: Prospective laboratory study. SETTING: Outpatient clinic with equipment in a standardized setting. SUBJECTS: Neonatal pneumotachometers. INTERVENTIONS: In standardized conditions, the tested pneumotachometer was calibrated at a flow rate of 3 L/min with 60% oxygen and was set in series with a closed spirometer system being used as a reference. Different air-flow levels (1-9 L/min) and oxygen concentrations (21-100%) were infused into the closed system with the pneumotachometer and spirometer. MEASUREMENTS AND MAIN RESULTS: The pneumotachometers were significantly affected by changing oxygen concentrations (p < .01) and increasing flow rates (p < .01), increasing the actually measured flow rate. Correction factors, developed by multiple regression analysis, significantly reduced the overall maximum errors of the pneumotachometers from -1.1 to 0.6 L/min to -0.5 to 0.4 L/min. CONCLUSIONS: The effects of changes in oxygen concentrations and flow rates on neonatal pneumotachometers could be considerably decreased by the use of correction factors such as were calculated in this study. This will preclude frequent calibration procedures with actual flow and oxygen levels during changes in experimental settings.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Correction factors for oxygen and flow-rate effects on neonatal Fleisch and Lilly pneumotachometers
- Date Crossref
- 01/04/2003
- Éditeur
- Ovid Technologies (Wolters Kluwer Health)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.