Abstract PD02-05: Comparison of fluorescence in situ hybridization (FISH) and dual-ISH (D-ISH) in the determination of HER2 status
Rattachement africain : us. Niveau de preuve : code pays fourni par la source.
Le résumé fourni par la source
Abstract Introduction: The appropriate assessment of HER2 amplification status is critical to determining which patients should receive HER2-directed therapy. An alternative to FISH is D-ISH (Ventana Medical Systems Tucson, Arizona) which utilizes chromogenic HER2 and chromosome 17 probes. Both FISH and D-ISH utilize formalin-fixed, paraffin-embedded (FFPE) human breast cancer tissue specimens. D-ISH has two principle advantages over FISH: it can be visualized using light microscopy and after analysis the specimens can be archived. Methods: We tested 250 samples with FISH and with D-ISH. Fifty specimens were controls: 25 cases with no HER2 or CEP17anomaly and 25 cases with HER2 amplification. There were also 200 test subjects of 50 cases each of chromosome 17 aneusomy, HER2 deletion (HER2/CEP17 ratio <0.8), HER2 duplication (HER2/CEP17 ratio 1.3–1.8), and equivocal results (HER2/CEP17 ratios between 1.8 and 2.2). FISH (Abbott Molecular) and D-ISH analysis were performed using methods described in the appropriate clinical kit package inserts, with the following exception: Four cases with CEP17 and HER2 co-amplification were also included and for FISH analysis these cases were reflexed to the use of chromosome 17 control probe (D17S122). Results: Four samples failed testing by D-ISH. There was a 63% (155/246) concordance between FISH and D-ISH by anomaly (aneusomy, duplication, deletion, etc.) and an 83% concordance by amplification status (non-amplified, equivocal, and amplified). D-ISH detected 18 (62%) of F-ISH-amplified cases. D-ISH resulted in lower estimates of HER2/CEP17 ratios than FISH, and many cases that were equivocal by FISH were normal by D-ISH. D-ISH did not detect HER2 amplification in any of the four HER2 and CEP17 co-amplified cases. Conclusions: We observed a lower concordance rate between FISH and D-ISH than the ∼95% rate in the D-ISH package insert. This lower concordance rate is likely a result of how this study was designed: it was biased toward common chromosome 17 anomalies and difficult specimens. Such specimens comprise about 60% of our current clinical HER2 biomarker practice. D-ISH may underestimate the HER2/CEP17 ratio, or FISH may overestimate this ratio. D-ISH also does not reliably detect HER2 amplification when the centromere of chromosome 17 is also co-amplifed. Compared to FISH, in this study cohort D-ISH had a significant false negative rate for the detection of HER2 amplification. Citation Information: Cancer Res 2012;72(24 Suppl):Abstract nr PD02-05.
Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.
Le contrôle bibliographique ouvert
DOI retrouvé dans Crossref DOI retrouvé ; titre concordant.
- Titre Crossref
- Abstract PD02-05: Comparison of fluorescence in situ hybridization (FISH) and dual-ISH (D-ISH) in the determination of HER2 status
- Date Crossref
- 01/12/2012
- Éditeur
- American Association for Cancer Research (AACR)
- Type
- journal-article
Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.
Les institutions déclarées
Une affiliation ne permet pas de déduire la nationalité d’un auteur.