Comparing and evaluating the efficacy of the TOR18FG Leeds test X-ray phantom for T-rays.

2Citations signalées, ce qui n’est pas une note de qualité
1Institutions déclarées
1Pays d’affiliation déclarés

Rattachement africain : au. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

The commercially available X-ray fluoroscopy quality assurance phantom, the Leeds test object TOR18FG, was found to be suitable to assess T-ray image quality in the range (0.1-0.4) THz at a depth of 0.5 cm. Previous to this only custom made phantoms, made especially for the T-ray region, assessed T-ray spatial resolution. However, if sub-wavelength techniques are used, the Leeds test phantom may be implemented to measure the T-ray systems spatial resolution, allowing us to directly compare X-ray and T-ray spatial resolution. The systems compared include a Gulmay Orthovoltage machine (X-ray), the On Board Imager (OBI) of a Varian linear accelerator (X-ray), a two-colour system (T-ray) and Terahertz Time Domain Spectroscopy (THz-TDS) system. X-rays were found to have a spatial resolution of 1.25 lp/mm using the On Board Imager of a Varian Linear Accelerator whilst T-rays imaged using a broadband source imaged through a spatial pinhole had a spatial resolution of 0.56 lp/mm. The TOR18FG background material was found to block, 90% and 99% of the broadband T-rays emitted from a THz-TDS photo-conductive emitter, at 0.4 THz and 0.53 THz respectively. Contrast sensitivity was found to be 3% for 25 cm × 25 cm X-ray field at 65 kV, whilst this value could not be established for T-rays using the TOR18FG. All contrast circles were found to be the same for T-rays i.e. all 40% at 0.1 THz. Images of the same leaf were taken with diagnostic X-rays and both broadband and continuous wave (CW) T-ray systems. T-rays proved superior in providing image contrast, for a hydrated leaf, over X-rays.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

La source scientifique ouverte est momentanément indisponible.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Terahertz technology and applicationsRadiation Detection and Scintillator TechnologiesDigital Radiography and Breast Imaging

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.