Keratometry obtained by corneal mapping versus the IOLMaster in the prediction of postoperative refraction in routine cataract surgery

8Citations signalées, ce qui n’est pas une note de qualité
5Institutions déclarées
2Pays d’affiliation déclarés

Rattachement africain : gb, ca. Niveau de preuve : code pays fourni par la source.

Le résumé fourni par la source

BACKGROUND: To establish whether simulated keratometry values obtained by corneal mapping (videokeratography) would provide a superior refractive outcome to those obtained by Zeiss IOLMaster (partial coherence interferometry) in routine cataract surgery. DESIGN: Prospective, non-randomized, single-surgeon study set at the The Royal United Hospital, Bath, UK, District General Hospital. PARTICIPANTS: Thirty-three patients undergoing routine cataract surgery in the absence of significant ocular comorbidity. METHODS: Conventional biometry was recorded using the Zeiss IOLMaster. Postoperative refraction was calculated using the SRK/T formula and the most appropriate power of lens implanted. Preoperative keratometry values were also obtained using Humphrey Instruments Atlas Version A6 corneal mapping. MAIN OUTCOME MEASURES: Achieved refraction was compared with predicted refraction for the two methods of keratometry after the A-constants were optimized to obtain a mean arithmetic error of zero dioptres for each device. RESULTS: The mean absolute prediction error was 0.39 dioptres (standard deviation 0.29) for IOLMaster and 0.48 dioptres (standard deviation 0.31) for corneal mapping (P = 0.0015). Keratometry readings between the devices were highly correlated by Spearman correlation (0.97). The Bland-Altman plot demonstrated close agreement between keratometers, with a bias of 0.0079 dioptres and 95% limits of agreement of -0.48-0.49 dioptres. CONCLUSIONS: The IOLMaster was superior to Humphrey Atlas A6 corneal mapping in the prediction of postoperative refraction. This difference could not have been predicted from the keratometry readings alone. When comparing biometry devices, close agreement between readings should not be considered a substitute for actual postoperative refraction data.

Ce résumé expose les affirmations des auteurs. BNTIC ne l’interprète pas comme une validation indépendante des résultats.

Le contrôle bibliographique ouvert

DOI retrouvé dans Crossref DOI retrouvé, mais le titre doit être comparé manuellement.

Titre Crossref
Keratometry obtained by corneal mapping <i>versus</i> the IOLMaster in the prediction of postoperative refraction in routine cataract surgery
Date Crossref
01/01/2013
Éditeur
Wiley
Type
journal-article

Ce recoupement confirme des métadonnées liées au DOI. Il ne confirme ni la méthode ni les conclusions de l’étude, et il ne compte pas comme une seconde source scientifique indépendante.

Les institutions déclarées

Une affiliation ne permet pas de déduire la nationalité d’un auteur.

Les sujets associés

Ophthalmology and Visual Impairment StudiesCorneal surgery and disordersIntraocular Surgery and Lenses

BNTIC News n’est pas le producteur de ces données. Les publications sont interrogées à la demande dans Crossref, OpenAIRE, DOAJ, Europe PMC, HAL, DataCite, AfricArXiv, ROR et la Banque mondiale, sans clé d’accès. OpenAlex reste optionnel. Aucun service payant n’est nécessaire et aucune donnée externe n’est enregistrée en base. Consulter les sources et leurs limites.